The Verification Report prints data for up to 7 captured curves on a graph based on real ear response. In addition, details regarding the hearing aid's configuration, audiogram, and adaptive features for the selected memory print on the report. If fitting binaurally, each hearing aid prints on its own page. However, if fitting CROS/BiCROS devices, only the receiver aid prints. The Report Header options selected via Reports Preferences determines the logo/text and patient information that prints at the top of the report. This report is only available if at least one curve was captured via Speech Mapping, Integrated Speech Mapping, or select Fitting or Advanced Tools screens during the current or retrieved session.
Click on the Read Device Settings or Load a Previous Session radio button and then click Start.
Click Advanced Tools from the Left Navigation Menu.

| Tip: To launch Integrated Speech Mapping, click Integrated Speech Mapping from the Speech Mapping screen. |

| Note: Only the captured curves checked View print on the report. |

| Note: Only the curves (e.g., Audiometric Thresholds, Soft Target/Response, etc.) checked print on the report. |
Click Print to print and/or preview the report.

Tip: Click on the Inspire Toolbar to print
and/or preview the report. |

| Note: Verification Report is only enabled if at least one curve was captured and checked View via Speech Mapping or select Fitting or Advanced Tools screens during the current or retrieved session and the current memory selected is not Disabled. |
| Note: If more than 7 captured curves were checked View via Speech Mapping or select Fitting or Advanced Tools screens, a message that only the 7 most recent curves will print on the report. Click Continue to print the 7 most recent curves or click Cancel to cancel the report and select the 7 desired curves to print. If real-time view Input + Gain was selected when the curve was captured, both an input and output curve were created, therefore, it's listed as 1 curve but is counted as 2. |
| Verification Report | Description |
| Report Header | Manufacturer's logo, company's logo, or customized text. Prints based on the Report Header selected via Reports Preferences. |
| Customized for | Patient's name. Only prints if Patient Name was selected via Reports Preferences. |
| DOB | Patient's date of birth. Only prints if Patient Date of Birth was selected via Reports Preferences. |
| Session | Date of the current or retrieved session. Only prints if Session Date / Date of Test was selected via Reports Preferences. |
Device/Audiogram |
|
| Right/Left Ear, Matrix, Acoustic Options, FW, Serial #, and Fitting Formula | Indicates the right/left hearing aid's product family, circuit, and style, firmware version, serial number, and the current configuration of the device's features. |
| Audiogram Graph | Office system, hearing aid, or In-situ air conduction, bone conduction, and UCL points. |
Memory and Current Output/Gain/Slope |
|
| Graph Legend | Identifies the curves, shaded regions, and captures on the graph. |
| Speech Mapping Captured Graph | Regardless of the graphical view currently selected, the fitting graph prints based on Real Ear Response output and Input Type: Speech. Only the curves (e.g., captures, Audiometric Thresholds, Soft Target/Response, etc.) checked via Speech Mapping or select Fitting or Advanced Tools screens print on the report. |
| Feedback Canceller, Directionality, Speech in Loud Noise, Speech in Noise/Speech and Noise/Noise, Quiet, Transients, Wind, Machine Noise, and/or Auto Music | Indicates the current setting of the memory's features. Only the features supported print. |