Speech Mapping allows you to illustrate to your patient how the hearing aid is processing speech or any other input in real-time by capturing curves and displaying them for easy understanding and verification of performance. It also allows you to provide counseling, fitting, and troubleshooting support during the initial fitting and follow-up fittings. A copy of up to 6 captured curves can be printed for reference. This feature is available for all product families and tiers except basic and economy.
Speech Mapping is only available during a live fitting. Curves can either be captured via Speech Mapping or any Fitting or Advanced Tools screen with a graphical view except Frequency Lowering, Hearing Aid Test, SoundPoint, and Verify Comfort. The options to capture, manage, pause/play, and select the desired views are available via both avenues. Once Speech Mapping is activated, the frequency response graph changes to display the real-time information of the hearing aid(s) via bars or lines for speech input.
Inspire provides numerous audio files to play (e.g., Airplane, Car Horn, Rain, etc. and samples of the various voice indicators) via Media Player, however, audio files and categories (e.g., Moderate Sounds, Speech, etc.) can be added and maintained via Audio Files Preferences. Prior to using Speech Mapping, it is important to calibrate your system speakers. See Speaker Calibration for more information.
Integrated Speech Mapping is an automated process of Speech Mapping that assists with verifying the hearing aid operation to assure good speech audibility and comfort for the fitting. It uses a calibrated input signal with a known spectral shape in order to give the most accurate possible measure of the hearing aid output in response to a known speech input. Integrated Speech Mapping first calibrates the hearing aid's input signal and then captures speech to verify the performance for Normal memory. This wizard can be launched via this screen or at the conclusion of Adult Auto Path.
Connect and select the desired hearing aid(s) to capture via Get Started.
Click on the Read Device Settings or Load a Previous Session radio
button and then click
.
Click Advanced Tools from the Left Navigation Menu.
Click Speech Mapping from the Left Navigation Menu or click Speech Mapping from select Fitting or Advanced Tools screens.
Note: Regardless
of the graphical view currently selected, Real
Ear
Response output
and Input Type: Speech always displays on the
fitting graph while Speech Mapping is activated. In addition,
Acoustic Options is hidden and replaced with a panel of Speech
Mapping buttons (e.g., ,
,
etc.). |
Tip: For
instructions on calibrating the speakers using the input monitor, click from the
Speech Mapping screen. Click X
in the upper right hand corner to close the instructions.
To launch Integrated
Speech Mapping, click |
| Tip: Click Tools from the Menu Bar and click Speech Mapping to activate. |
Click on the specific memory to capture if applicable.
| Note: If capturing Honor, Pride, or Pride IIC products, this feature is not available for Telecoil, Autocoil, or Loop memories and for the Music memory, the standard fitting graph, rather than the music equalizer graph, displays. |
Inspire automatically defaults to a binaural
or
view if applicable.
To capture only the right or left hearing aid, click
or
,
or
or
.
| Note: If fitting CROS/BiCROS devices, only the receiver aid can be captured, therefore, the side cannot be selected. |
The table below provides the recommended stimuli, display, capture, and adjustments based on the patient's issue. Please refer to this table when completing steps 8-14.
| Category | Patient's Issue | Recommended Stimuli | Recommended Display and Capture | Recommended Adjustments |
| Loudness | Overall
Too Loud
|
|
|
|
Overall
Too Soft
|
|
|
|
|
Loudness
Comfort
|
|
|
|
|
| Feedback | Hearing
Aids
|
|
|
|
| Sound Quality | Noisy
|
|
|
|
Pumping
|
|
|
|
|
Shutting
Down
|
|
|
|
|
Sounds
|
|
|
|
|
| Noise | Difficulty
|
|
|
|
Too
Much Noise
|
|
|
|
Check or uncheck the curves to display or hide
on the fitting graph in the Display
section on the Speech Mapping screen or click Display
or
.
| Note: If None is selected for any target level via Target Preferences, that option is disabled. For the Music memory in Honor, Pride, or Pride IIC products, regardless of the curves selected, only the Audiometric Thresholds display (if checked). |
Select the real-time views to display on the
fitting graph from the drop down lists on the Speech Mapping
screen or click View
or
.
| Note: The default views are Input + Gain and Bars. Input is represented in gold bars or lines and Gain/Output is represented in red or blue. If Bars is selected, peak-level indicators display for each band and adjust as the real-time display increases or decreases. If Frequency Lowering is enabled in the selected memory, Frequency Lowering precedes the word Input and if selected, displays the input to multichannel expansion compression output. |
Start the stimuli (e.g., talking, clapping, playing an audio file via Media Player, etc.).
Click
or
to capture the
average response
over time. Click
or
to capture a single response and a
curve of the same color displays on the fitting graph.
Tip: The
average response
capture automatically stops recording after 10 seconds. To
record less than 10 seconds, click
or
to stop
the recording at any time. Once the average
response capture is complete, a curve encased in a
shaded region of the same color displays on the fitting graph. |
Manage the curve in the Captures
section on the Speech Mapping screen or click Captures
or
.
Note: Inspire
automatically assigns the curve color in a pre-determined
sequence (e.g., black,
green, etc.), therefore,
based on the number of curves captured previously, determines
the color assigned. However, any curve color can be changed
by clicking on the color swatch (e.g., ,
, etc.). |
| Field Name | Description | Details |
| View | Denotes whether or not to display the captured curve on the fitting graph. | To display all captures, click View All. |
, , or
![]() |
Icon representing the ear side and view captured (Input, Input + Gain, or Output). | If a binaural speech curve was captured via Integrated Speech Mapping, the right and left captures may be listed separately. |
| Curve Label | Describes the captured curve. The color, memory (in parenthesis), and time of the capture automatically defaults. | Select the text to change, if desired, and type the new label. This label prints on the Verification Report. |
| Color Swatch | Changes the color of the captured curve. | Click on the
color swatch (e.g., , , etc.) and a color
palette displays. Click on the desired color. |
| Delete | Deletes the captured curve. | Click
to delete the captured curve. To remove all captured curves,
click Delete All. To reverse
any deletions, see Undo. |
| Note: The curve label is highlighted and displays in a tool tip when pointing to the captured curve on the fitting graph. |
Click
to pause the real-time display or to make adjustments.
| Note: Adjustments can only be made graphically by dragging and dropping points on the frequency response curve on the Speech Mapping screen when the real-time display is paused. Adjustments can be made in any view graphically or via buttons/sliders on select Fitting screens when the real-time display is playing or paused. |
Tip: The
button changes to .
Click
to resume the real-time display. |
Make adjustments if necessary.
To stop Speech Mapping, click X
to the right of
from select Fitting or Advanced
Tools screens
or click
.
| Tip: Click Tools from the Menu Bar and click Stop Speech Mapping to deactivate. |
Click
on the Inspire Toolbar to print and/or preview
the Verification
Report if desired.
| Tip: Click File from the Menu Bar and click Print to print and/or preview the report. |