| Field Name |
Description |
Details |
| Soft |
Selects
the default soft target level. |
The
levels available are None,
40,
45,
50, and
55. The
default value is 50 dB SPL.
If None
is selected, the soft target and response
curves will not display. |
| Moderate |
Selects
the default moderate target level. |
The
levels available are None,
60,
65,
and 70.
The default value is 65
dB SPL. If None
is selected, the moderate target and response
curves will not display. |
| Loud |
Selects
the default loud target level. |
The
levels available are None,
75, 80,
85,
90, and
95. The
default value is 80
dB SPL. If None
is selected, the loud target and response
curves will not display. |
| Target
Input Stimulus |
Selects
the desired input stimulus for the fitting formula target
calculation. |
The options
available are Speech
and Sinusoidal.
The default is Speech. |
| Binaural
Correction |
Selects the desired target correction
value used in binaural fittings to compensate for the
binaural summation effect obtained when fitting both ears.
The fitting formula target for each ear will be reduced
by the dB value selected. |
The options available are 0,
3, and 5 dB. The default value
is 3 dB. |
| Air-bone
Gap Correction |
Selects the desired target correction
value to be used when the audiogram indicates an air-bone
gap, or a difference between the thresholds obtained by
air conduction and bone conduction. The fitting formula
target values will be increased by the percentage of the
air-bone gap selected. |
The options available are 0,
25, 50,
75, and 100%. The default air-bone
gap correction is 25%. |
Fitting
Formulas |
| Non-Linear |
Selects the non-linear fitting formulas
that will be available in the drop down list from any
Fitting screen. |
The non-linear fitting formulas
available are e-STAT,
NAL-NL2, NAL-NL1, and DSL v5. By default,
all fitting formulas are selected. |
| DSL Defaults
(Button) |
Identifies the default settings exclusively
for the DSL v5 fitting formula. |
See DSL
Defaults for more information. |
| Linear |
Selects the linear fitting formulas that
will be available in the drop down list from most Fitting
and Advanced Tools screens. |
The linear fitting formulas
available are NAL-R,
Berger, and WFA.
By default, only NAL-R
is selected. |
| Select Default Formula |
Selects
the default fitting formula for new hearing aids or when
simulating hearing aids. |
The
fitting formulas available are based on the non-linear
and linear fitting formulas selected. The default fitting
formula is e-STAT. |
Experience
Level |
| Apply
default at Best Fit and
Target Match |
Denotes
whether or not to apply an automatic gain offset within
the fitting. |
By default,
this gain offset is automatically applied. |
| Experience Level |
Selects
the default gain offset to apply within the fitting. |
This
field is only enabled if Apply
default at Best Fit and Target Match is checked.
The options available are Novice
(1), Familiar
(2), Experienced
(3), and Transition
(T). The default level is Experienced (3). |
| Show Auto Experience Manager
final settings when the feature is enabled |
Denotes
whether or not to display the Auto Experience Manager's
final gain settings on the fitting graph when this feature
is enabled in the hearing aid connected or simulated. |
If
selected, which is checked by default, the Auto Experience
Manager's final gain settings will display on the
fitting graph represented as dotted lines. If not selected,
the fitting formula's targets will display on
the fitting graph represented as thin lines. |
Best
Fit Options |
| Perform
on fitting formula change |
Denotes whether or not
to perform a Best Fit every time a different fitting formula
is selected. |
The default is to perform
a Best Fit whenever the fitting formula is changed. |