Patient Tools for Evolv AI, Livio Edge AI, Livio AI, and Livio Products

Patient Tools performs a diagnostic test of the components in your patient's hearing aids (Self Check), conducts TeleHear Remote Services, and/or enables/disables Auto Fall Alert.

  1. Connect and select the desired hearing aid(s) to test via Get Started.
  2. Click on the Read Device Settings radio button and then click Start.
  3. Click Patient Tools.
    Tip:  This screen can also be accessed via Fitting Summary by clicking Self Check or Auto Fall Alert under Status.
  4. Click to expand the desired section if not currently expanded.
    Note:  Only one section can be expanded at a time. Therefore, if Self Check is currently expanded and then Auto Fall Alert is expanded, Self Check will automatically collapse.

Self CheckSelf Check

 

Self Check performs a diagnostic test of the components in your patient's hearing aids. Once a baseline is measured for the receiver, microphone, circuit, and heart rate sensor (if applicable), a diagnostic test can be run via Inspire or the Thrive Hearing Control app. If the results determine that a component is not performing to baseline or fails the test, the hearing aid may need to be cleaned or sent in for repair.

 

Self Check BaselineSelf Check Baseline

 

The Self Check Baseline is performed during the initial fitting for Livio AI and Livio products, while the hearing aid is connected to Inspire. The measurement must be performed in a quiet environment, while the devices lay on a flat hard surface with the receiver facing up. The baseline is obtained by the hearing professional and it is assumed that the device is performing optimally. The Self Check Baseline measures and records the status of the microphone, receiver, circuit, and heart rate sensor.

 

It is recommended that additional baseline measurements are performed after any factory repairs, as baseline data is cleared by the factory.

 

For Evolv AI and Livio Edge AI products, the Self Check Baseline is set by the factory. However, if there are changes to the acoustic options, the baseline threshold is automatically updated.

 

Self Check DiagnosisSelf Check Diagnosis

 

Subsequent Self Check measurements, referred to as Self Check Diagnosis, can be performed by the patient in the field or by the professional in their office. Self Check Diagnosis measures are compared to the baseline to determine status.

 

After each successful completion, the hearing aids report status to the patient or professional.

 

The most recently collected Self Check data stored in the hearing aid can be viewed in Inspire. Test results saved to a previous fitting session can be viewed if retrieved.

Note:  If viewing binaurally, the baseline or test is run in the right aid first and then the left aid.
  1. Inspire automatically defaults to a binaural Select Binaural view if applicable. To test only the right or left hearing aid, click Select Right Only or Select Left Only.
    Note: If testing CROS/BiCROS devices, only the receiver aid can be tested, therefore, the side cannot be selected.
  2. Place the hearing aid(s) on a flat hard surface with the receiver facing up and reduce all sources of noise.
    Note:  It's recommended that Self Check be performed in a quiet environment (baseline ≤ 60 dB SPL; diagnostic test ≤ 85 dB SPL) with no speech/talking.
  3. If a baseline has not been set, click Set Baseline to start the measurement and a brief, robust sound plays in each hearing aid.
    Note:  This button is only enabled during a live fitting for Livio AI and Livio products. If the baseline was successful for a component, displays. If unsuccessful, displays.
    Tip:  The Set Baseline button changes to Cancel during the measurement. Click Cancel to cancel the measurement.
    Note:  If the baseline was set previously, it can be run again if desired.
  4. Click Begin Test to run the diagnostic test and a brief, robust sound plays in each hearing aid.
    Note:  This button is only enabled during a live fitting if the baseline was set previously.
    Tip:  The Begin Test button changes to Cancel during the test. Click Cancel to cancel the test.
    Tip:  The patient can also run the test in the Thrive Hearing Control app by tapping Run Diagnostic via Settings>Self Check.
  5. Review the test results and follow any recommended steps if necessary.
    Status Inspire Icon Thrive Icon Description

    Pass

    The device is performing to baseline or close to baseline.

    Warning

    The device component is not performing to baseline and should be checked for debris. Clean the aid, change the wax guard, etc. and run the diagnostic test again.

    Fail

    The device component or components may need repair. Before sending in for repair, clean the aid and run the diagnostic test again.

QuickTIP

 

TeleHear Remote Services

 

Auto Fall Alert