SurfLink Wireless Diagnostics

SurfLink Wireless Diagnostics assists in troubleshooting issues with your wireless connection and provides solutions to the identified problems. Three components of your wireless connection are checked: programmer status, environmental status, and wireless communications status. Each component contains several subtests and all subtests within a component must pass before the next component can be checked. If a subtest within a component fails, wireless diagnostics is aborted.

 

The quality of the wireless environment in your fitting area can also be assessed. Inspire logs any wireless interference in your local environment and displays it on a real-time graph. These diagnostic results can be emailed to Inspire Software Support for assistance.

  1. Click Tools from the Menu Bar.
  2. Click SurfLink Wireless Diagnostics.
  3. Click Start and a progress icon indicates the overall status of completion.
  4. When the component Checking Wireless Communications Status is reached, the Devices Detected pop-up displays to select a wireless device. Click the checkbox to the left of the desired hearing aid(s) to test.
  5. Click OK.
  6. When instructed, place the hearing aid(s) 1 meter from the SurfLink programmer.
  7. Click OK.
  8. When instructed, place the hearing aid(s) 3 meters from the SurfLink programmer.
  9. Click OK.
  10. Click Send Files to send diagnostic results to Inspire Software Support if additional assistance is required.
  11. Enter email information.
  12. Click Save to Desktop to save the log files to a .zip file on the desktop or click Send Email to send the log files to Inspire Software Support.
  13. Once the files are saved or sent, a message displays. Click OK.
  14. Click Advanced to assess any wireless interference if applicable or click Close to close wireless diagnostics.
  15. Select graph settings.
  16. If interference is detected, turn off any devices (e.g., cell phone, SurfLink Media, Plantronics CS50 headset, paging system, etc.) that may be interfering with the SurfLink Programmer.
  17. Click Close if the interference is resolved or click Send Files to send diagnostic results to Inspire Software Support if the interference is not resolved.
  18. Click Close to close wireless diagnostics.